(Reference Quantity: Full Range Deflection = FRD)
•5 Measurement Ranges
0-10 ppm Propane
0-100 ppm Propane
0-1.000 ppm Propane
0-10.000 ppm Propane
0-100.000 ppm Propane
•Linearity:+/- 2 % MBE (up to Range 4)
•Oxygen Cross Sensitivity:< +/- 2 % FRD (with He/H2 Fuel)
•Limit of Detection:< 2 % FRD
•Zero Driftt:< 2 % FRD/week
•Span Drift:< 4 % FRD/week
•Response Time90:approx. 1-2 s
•Sample Flow Rate:approx. 90 l/h
•Analytical Temperature:up to 200°C
•Signal Output:0-10 V and/or 0/4-20 mA
•Status Contacts:Concentration-Alarm、Range Identification、Flame Out Indication、Sample Flow Alarm
•Dimensions:360 x 200 x 500 mm (WxHxD)
•Weight:approx. 20 kg
•Ambient Temperature:+5 ... +35 °C
•Main Power:230 V / 50 Hz or 115 V / 60(50) Hz / 450 Watts